Suture Anchors and How Repairs Are Held to Bone

102 citationsUpdated Oct 2026
Illustration: shoulder

For patients: a plain-language version of this topic is available. See the patient guide.

Overview

Suture anchors and tacks serve as the primary interface for securing soft-tissue repairs to bone, with their successful use governed by biological and biomechanical factors [2]. The field has undergone a major shift from metallic to bioabsorbable anchors, driven by higher complication rates associated with metallic devices and lower rates with bioabsorbable alternatives [35]. All-suture anchors possess similar or better mechanical properties than regular anchors, and their low-profile design preserves bone tissue [5]. While various bone changes are observed after implantation of an all-suture anchor, these changes are not clinically relevant [1]. The 2.6-mm all-suture anchor mitigates risks and difficulties associated with other fixation devices while providing ease and efficiency for the surgeon [4].

Complications related to suture anchors and tacks can be categorized as technique-related or device-related issues [9]. Prevention of these complications depends on the surgeon's familiarity with the devices and knowledge of their indications and limitations [9]. Careful attention to proper anchor insertion techniques can limit the potential for complications [6], while newer materials may address concerns regarding biocompatibility and material strength in bioabsorbable anchors [6]. Routine radiographic follow-up is recommended after the use of metallic anchors to ensure identification of early failure by anchor pullout [11].

Outcomes vary by specific application and technique. Suture anchors simplify the open Bankart procedure, with no specific complications seen as a result of the modified open Bankart technique using suture anchors [12]. Good long-term results were observed after arthroscopic Bankart repair in patients older than 20 years with 3 or more suture anchors used [21]. Arthroscopic transosseous superior capsular reconstruction outcomes are not inferior to those of anchor-based techniques, involving less suture tangling, more precise suture pass points, and lower costs [8]. Similarly, arthroscopic transosseous rotator cuff repair using a bone tunneling device facilitates robust and reproducible repairs and substantially reduces procedural costs compared to suture anchor techniques [37]. Transosseous suture fixation is not necessarily a "weak link" of repairs [25].

Anatomy & Pathophysiology

Bony Anatomy

The proximal humerus comprises four main parts: the humeral head, greater tuberosity (GT), lesser tuberosity (LT), and humeral shaft [63]. The articular head of the humerus is spherical with a diameter of 37 to 57 mm [63] and a radius of curvature of approximately 25 mm [75]. The articular surface is essentially spherical, with an arc of approximately 160 degrees covered by articular cartilage [75]. The most superior portion of the articular surface averages 8 mm above the greater tuberosity [63], while the superior margin of the humeral head articular surface is normally superior to the top of the greater tuberosity by 8 to 10 mm [75]. Humeral version averages 29.8 degrees, with a range of 10 to 55 degrees [63], and the humeral head is retroverted an average of 30 degrees [64] relative to the transepicondylar axis of the humerus [77]. Other sources report an average retroversion of 19° [66]. The humeral head is inclined approximately 130 degrees with respect to the humeral shaft [63], and the neck-shaft angle of the proximal humerus measures an average of 135 degrees [64]. Another measurement reports the average neck-shaft angle as 45 degrees (±5 degrees), with a range of 30 to 50 degrees [75]. The greater tuberosity serves as the attachment site for the supraspinatus, infraspinatus, and teres minor tendons of the rotator cuff [63], while the lesser tuberosity serves as the attachment site for the subscapularis tendon [63]. The distance from the lateral base of the coracoid process to the lateral margin of the greater tuberosity is called the lateral humeral offset [75].

The glenoid is a convex structure of shallow depth shaped like an inverted pear [63] and is a shallow socket, approximately one third the size of the humeral head [64]. The glenoid articular surface radius of curvature is 2 to 3 mm larger than that of the humeral head [75]. The subchondral bone of the glenoid is relatively flat, with articular concavity augmented by cartilage and a circumferential labrum [66]. The glenoid is retroverted approximately 5 degrees relative to the scapular body [66, 77]. The scapula is attached to the axial skeleton by the acromioclavicular (AC) and sternoclavicular (SC) joints [65]. The glenoid is connected with the flat body of the scapula by the scapular neck [65]. The coracoid process curves forwards from the superior surface of the scapular neck [65], and the acromion is a flattened bony process that curves forwards from the scapular spine [65]. The highest concentration of bony mass in the scapula is located in the glenoid, the scapular neck, and the lateral border of the scapular body [65]. Two bony pillars transmit compressive forces from the glenoid fossa: the lateral pillar and the spinal pillar [65]. The weakest bone in the scapula is located primarily in the central part of the infraspinous fossa [65], and the weakest area of the circumference of the biomechanical body of the scapula is the spinomedial angle [65].

Soft Tissue Anatomy

The rotator cuff is a sheet of conjoined tendons closely applied over the shoulder capsule, inserting mainly into the greater tuberosity of the humerus [71]. The subscapularis tendon inserts into the lesser tuberosity [71]. The coracoacromial arch is formed by the acromion process posterosuperiorly, the coracoid process anteriorly, and the coracoacromial ligament joining them [71]. The subacromial bursa separates the rotator cuff tendons from the coracoacromial arch, allowing them to glide [71]. The subscapular bursa lies between the subscapularis tendon and the neck of the scapula [67] and communicates with the joint cavity between the superior and middle glenohumeral ligaments [67]. The rotator interval is defined medially by the base of the coracoid, superiorly by the supraspinatus tendon, and inferiorly by the subscapularis tendon [66]. The rotator interval contains the coracohumeral (CH) ligament, the superior glenohumeral ligament (SGHL), and the intra-articular portion of the long head of the biceps tendon [66].

The tendons of the infraspinatus and supraspinatus muscles join approximately 15 mm proximal to their insertion [76]. The supraspinatus and subscapularis tendons join as a sheath that surrounds the biceps tendon at the entrance of the bicipital groove [76]. The coracohumeral ligament is a thick band of fibrous tissue extending from the coracoid process along the surface of the capsule to the tuberosities between the supraspinatus and subscapularis tendons [76]. The glenohumeral ligaments provide static restraint based on arm position. The superior glenohumeral ligament is the primary restraint to inferior humeral subluxation in 0 degrees of abduction [76] and the primary static restraint against anterior translation with the arm at the side [66]. It also restrains external rotation and inferior translation of the adducted or slightly abducted arm [77]. The middle glenohumeral ligament limits external rotation when the arm is in the lower and middle ranges of abduction [76] and is a primary static restraint against anterior translation with the arm in external rotation and 45° of abduction [66]. It restrains anterior translation with the arm abducted to 45 degrees [77]. The inferior glenohumeral ligament is composed of an anterior band, a posterior band, and a thinner intervening axillary pouch, creating a hammock-type sling [76]. The anteroinferior glenohumeral ligament complex is the main stabilizer to anterior and posterior stresses when the shoulder is abducted 45 degrees or more [76]. The anterior band of the inferior glenohumeral ligament is a primary static restraint against anterior-inferior dislocation of the glenohumeral joint in 90° of abduction and external rotation [66] and restrains anterior and inferior translation with the arm externally rotated and abducted to 90 degrees [77]. The posterior band of the inferior glenohumeral ligament is a primary static restraint against posterior-inferior translation in internal rotation and adduction [66] and restrains posterior and inferior translation with the arm internally rotated and abducted to 90 degrees [77]. The coracohumeral ligament restricts external rotation in adduction and is a static restraint to inferior and posterior translation in adduction and external rotation [66].

Vascular Supply

The proximal humerus receives its blood supply from the anterior and posterior humeral circumflex branches from the third division of the axillary artery [63]. The anterior humeral circumflex artery provides vascular inflow to the humeral head by way of its terminal anterolateral branch known as the artery of Laing (arcuate artery) [63]. The ascending branch of the anterior humeral circumflex artery enters the humeral head at the interface of the bicipital groove and greater tuberosity [63]. The major blood supply to the humeral head is through the ascending branch of the anterior humeral circumflex artery, which penetrates the head at the bicipital groove and becomes the arcuate artery [64]. The anterolateral ascending branch of the anterior humeral circumflex artery provides the primary blood supply to the humeral head [66]. The terminal intraosseous portion of the anterior humeral circumflex artery enters at the proximal aspect of the intertubercular groove as the arcuate artery [66]. Quantitative assessment has shown that 64% of the humeral head blood supply arises from the posterior humeral circumflex artery [70].

Pathophysiology of Failure

The bony anatomy of the glenohumeral joint contributes little to stability, having been compared with a golf ball on a tee [76]. Most stability of the shoulder is provided by surrounding muscles and ligaments due to the lack of bony constraint [76]. The glenohumeral joint depends on static and dynamic stabilizers for movement and stability, especially the rotator cuff [75]. The rotator cuff stabilizes the glenohumeral joint while allowing greater freedom of motion and fixes the fulcrum of the upper extremity against which the deltoid can contract [75]. The rotator cuff muscles serve as depressors of the humeral head to allow the deltoid to efficiently abduct the humerus [64]. The infraspinatus and teres minor are external rotators, while the subscapularis is an internal rotator of the humerus [64]. The most important force couple for shoulder stability involves the subscapularis and posterior rotator cuff, providing a compressive force that centers the humeral head in the glenoid cavity [76].

The glenoid labrum increases the depth of the socket by 50% around the humeral head [76] and provides a bumper to translation [77]. The glenoid articular surface and labrum combine to create a socket approximately 9 mm deep in the superoinferior direction and 5 mm deep in the anteroposterior direction [76]. Adding the glenoid labrum increases the glenoid surface to 75% of the humeral head vertically and 57% horizontally [76].

First- and second-generation repair techniques fail at a rate in excess of 30% primarily because of sutures pulling through tendons or tendons failing to heal at the tendon-bone junction [106]. The healing zone at the tendon-bone junction is no more substantive than fibrovascular scar [106]. The goals of rotator cuff repair are to keep the tendon "long enough" and "good enough" to achieve pain reduction and improvements in strength and function [106]. In revision rotator cuff repairs, modes of failure include tendon pulling through suture, tear in a new location, suture anchors pulling out of bone, and suture breakage or suture knot [23]. In a study of 22 revision rotator cuff repairs, only 1 case was associated with a loose suture anchor [23]. Osteoporosis in the proximal humerus can result in poor fixation of the anchor to bone [23]. Suture anchors are contraindicated in elderly patients with osteoporotic bone according to several authors based on ex vivo investigations [23].

Biomechanical comparisons of repair configurations show that the biomechanical properties of a double-row rotator cuff repair are superior to a single-row repair [104]. However, there is no evidence that double-row fixation results in better functional outcomes compared to single-row fixation [113]. Rotator cuff tendons repaired via the knotted TOE technique display superior time-zero biomechanical properties, including greater ultimate load to failure, compared with rotator cuffs repaired via the knotless TOE technique [121]. There was conflicting evidence supporting knotless or knotted anchors in biomechanical studies for labral repair [130]. Operative times may be reduced with the use of knotless anchors for labral repair [130]. Bone fixation seems to be improved with knotless anchor systems [3]. The development of knotless suture anchors has allowed for the expansion of their use with elimination of the side effects caused by suture knots without impact on biomechanical strength [52]. By eliminating suture knots, postoperative trochanteric pain is reduced in abductor tendon repair [52]. The knotless, low-profile design eliminates a prominent knot stack superiorly and decreases the risk of stress riser in the acromion and distal clavicle compared to larger, solid body anchors [126].

Complications related to hardware include cartilage damage caused by migration of suture anchors or mini screws, which might progress and cause symptoms in the future even after the loosened anchor or screw is removed [31]. Shoulder function improved in all patients after the loosened anchor or screw was removed [31]. All-suture anchors with various deployment configurations produced different tunnel enlargements, with the diameter more pronounced at the inferior region of the anterior glenoid [131]. Ensuring proper anchor angle is necessary to avoid damage to the cartilage during knotless anchor placement [112]. The TOAK technique eliminates the potential risk for loose hardware in the subacromial space [17].

Clinical outcomes for various techniques include the following. The double-row, transosseous-equivalent suture-bridge repair facilitates the achievement of pain reduction and restoration of shoulder functionality [44]. The knotless double-row technique for reinforcing anterosuperior medium-to-large rotator cuff repairs aims to maximize footprint coverage, increase tendon thickness and effective length, and restore superior stability of the glenohumeral joint [128]. Significant improvements were observed across pain, range of motion, strength, and functional outcomes in a prospective cohort of 1000 consecutive cases of undersurface rotator cuff repair using the Opus knotless system [125]. The retear rate for undersurface rotator cuff repair using the Opus knotless system was 1.4% [125]. At 24 weeks, forward flexion and shoulder strength in abduction, adduction, external rotation, internal rotation, and lift-off improved compared with preoperative levels following arthroscopic stabilization using PEEK knotless anchors [120]. Current level II to IV clinical and biomechanical evidence supports knotless arthroscopic Bankart repair as an effective procedure with excellent functional outcomes and low rates of complication and reoperation [117]. The augmentation of large to massive, posterosuperior rotator cuff tears with human dermal allograft patch resulted in improved shoulder function and greater range of motion compared with the nonaugmented group [116].

Specific technical applications include the following. The Knotless FiberTak is equipped with a shuttling suture mechanism that allows the repair suture to be passed through its locking mechanism [53]. The knotless tensionable anchor technique allows for the desired tension to be given to the construct by pulling the SutureTape shuttle through the splice locking mechanism [53]. The zip-tie technique allows for greater control over tensioning while preserving joint access throughout the repair of multidirectional instability [112]. Premature capsular tightening in conventional approaches can create unnecessary tension, complicating subsequent anchor placement and reducing surgical flexibility [112]. The inability to fine-tune tension at the end of a procedure may lead to suboptimal capsular volume reduction [112]. The risk of shoulder loosening at 5-8 years postoperatively is a potential pitfall of the zip-tie technique for multidirectional instability [112]. The "Suture Staple" knotless anchor technique is suitable for tibial tubercle avulsion fractures where the fragment is too small for screw fixation [52]. The metal-free autograft bone block suture tape cerclage technique resulted in a 4º decrease in external rotation at the side and a 3º decrease in external rotation in 90º of abduction [59]. The arthroscopic transosseous equivalent technique for bony Bankart repair is simple and readily applied in any arthroscopic shoulder surgery practice [60]. The intra-articular subscapularis repair using the transtendon lasso-loop technique with an all-suture anchor expands the surgical armamentarium for managing complex subscapularis pathology [111]. The tensionable knotless tendon-sparing biceps rerouting technique is reproducible, minimally invasive, and suitable for degenerative or inflamed tendons [114]. The fixation of the proximal hamstring tendon using an all-suture tensionable knotless technique combines the biomechanical strength of using multiple smaller anchors with knotless fixation [127]. The single medial-row anchor with biceps tenodesis in a transosseous double-row construct is a promising technique for the repair of massive rotator cuff tendon tears due to technical efficiency and strong biomechanical construct [136]. The double-barrel knot is a new arthroscopic sliding knot for soft tissue reconstruction and repair procedures in arthroscopic and open shoulder surgery [138].

Classification

Complication Categories: Suture anchor complications are categorized as either technique-related or device-related issues [9]. Careful attention to proper anchor insertion techniques can limit the potential for complications with bioabsorbable anchors [6]. Newer materials for bioabsorbable anchors may address concerns of biocompatibility and material strength [6]. The new resorbable suture anchor offers the possibility of greater biocompatibility than currently available absorbable suture anchors [55].

Postage Stamp Fractures: Postage stamp fractures are frequently reported in patients who are male, age 25 years or younger, and participants in sporting activities [10]. These fractures are also frequently reported in cases initially stabilized with 3 or more anchors or conventional knot-tying anchors [10]. Additionally, postage stamp fractures are frequently reported in cases that experienced osteolysis around anchor sites [10].

Modes of Failure: Suture anchors pulling out of bone is one of the identified modes of failure for rotator cuff repair at revision surgery [23]. Tendon pulling through suture is one of the identified modes of failure for rotator cuff repair at revision surgery [23]. A tear in a new location is one of the identified modes of failure for rotator cuff repair at revision surgery [23]. Suture breakage or suture knot is one of the identified modes of failure for rotator cuff repair at revision surgery [23]. In a study of 22 revision rotator cuff repairs, only 1 was associated with a loose suture anchor [23]. The loose suture anchor in the revision study may have been the result of poor placement rather than the anchor pulling out [23].

Perianchor Bone Reaction: The perianchor bone reaction grade 0 was observed in 70.1% of patients with all-suture-type anchors [122]. The perianchor bone reaction grade 1 was observed in 21.2% of patients with all-suture-type anchors [122]. The proportion of patients with perianchor bone cysts (grade 2 or above) was 8.8% for all-suture-type anchors [122]. The perianchor bone reaction grade 0 was observed in 64.9% of patients with bioabsorbable-type anchors [122]. The perianchor bone reaction grade 1 was observed in 19.4% of patients with bioabsorbable-type anchors [122]. The proportion of patients with perianchor bone cysts (grade 2 or above) was 16.7% for bioabsorbable-type anchors [122]. The perianchor bone reaction grade 0 was observed in 72.5% of patients with PEEK-type anchors [122]. The perianchor bone reaction grade 1 was observed in 15.0% of patients with PEEK-type anchors [122]. The proportion of patients with perianchor bone cysts (grade 2 or above) was 12.5% for PEEK-type anchors [122]. No significant difference in perianchor bone reactions was observed among the different anchor types [122]. Various bone changes are seen after implantation of an all-suture anchor, but these are not clinically relevant [1].

Other Considerations: The bearing area of a suture anchor can be used to predict the pullout strength independent of design in a soft-bone model [15]. All-suture anchors have similar or better mechanical properties than regular anchors [5]. All-suture anchors have a low-profile design that preserves bone tissue [5]. The Micro bone suture anchor provides a stronger tendon to bone repair than the pullout button or the Mini anchor in human cadaveric hands [20]. Suture anchor fixation is stronger than screw fixation immediately after surgery for humeral greater tuberosity fractures [33]. The TOAK technique is a biomechanically strong and secure cuff fixation that eliminates the potential risk for loose hardware in the subacromial space [17]. No single technique is overwhelmingly superior for managing intraoperative anchor pullout during arthroscopic rotator cuff repair [22]. In a study of 342 patients, 21 (6%) subsequently underwent a revision rotator cuff repair by the same surgeon [23]. In a study of 342 patients, 1 patient had a second revision surgery [23]. A greater number of suture anchors results in greater strength of the repair [28]. The mean number of suture anchors varied among studies and within repair groups in the same study [28]. The repair technique (knotted suture-bridging or knotless tape-bridging) did not have a significant effect on the final outcome at a mean of 2.9 years [16]. Bone fixation seems to be improved with the knotless anchor systems in latissimus dorsi transfer fixation [3]. The small, 2.6-mm all-suture anchor mitigates risks and difficulties associated with other fixation devices while providing ease and efficiency for the surgeon [4]. The knotless double-loaded suture anchor technique for bony Bankart lesion fixation requires only 2 suture anchors, fewer than for double-row repair [7]. The knotless double-loaded suture anchor technique allows slow, controlled compression of the fracture during reduction [7]. The outcomes of arthroscopic transosseous superior capsular reconstruction are not inferior to those of anchor-based techniques [8]. Arthroscopic transosseous superior capsular reconstruction involves more precise suture pass points than anchor-based techniques [8]. The suture anchors were found to simplify the procedure in open Bankart operations [12]. No specific complications as a result of the modified technique using suture anchors were seen in open Bankart operations [12]. The arthroscopic retro all-suture anchor technique for subscapularis repair avoids a difficult trajectory and large-diameter bone punches during anchor insertion [14]. The arthroscopic retro all-suture anchor technique for subscapularis repair provides a reproducible and secure fixation [14]. A possible large defect is a disadvantage if the 4.0-mm-size knotless all-suture anchor were to pull out [26]. There is no clear differentiator between the looped and non-looped end of the passing suture in the joint when using a knotless all-suture anchor with suture tape [26]. The suture dyeing method is particularly useful in large tears with multiple anchors or when suture limbs are entangled [36]. The transosseous rotator cuff repair system using a bone tunneling device facilitates robust and reproducible repairs [37]. The transosseous rotator cuff repair system using a bone tunneling device substantially reduces procedural costs compared to suture anchor techniques [37]. The all-suture tape cerclage technique for AC joint separation fixation was designed to minimize the risk of complications inherent to currently popular procedures [102]. The stump classification was significantly correlated with retear in the suture-bridge and double-row repair techniques [105]. The dermal tuberoplasty technique using self-punching, knotless fixation simplifies graft delivery and suture management [108]. The suture staple knotless anchor technique for tibial tubercle avulsion fractures is based on the technique described by Ratner et al. for capsulotenodesis [52]. Knotless suture anchors allow for the expansion of their use with elimination of the side effects caused by suture knots without impact on biomechanical strength [52]. Lacheta et al. found similar biomechanical strength for first failure load, ultimate load, and stiffness between knotless all-suture anchors and knotted all-suture anchors [52]. Hartigan et al. described that eliminating suture knots in abductor tendon repair reduces postoperative trochanteric pain [52]. The goals of the review on suture anchors and tacks are to review the biological and biomechanical factors that influence the successful evaluation and use of suture anchors and tacks [2]. The repair technique using a combination of suture anchor and miniscrew can be applied in any similar case, regardless of suture anchor type [19]. Future studies are required to validate the biomechanical principles, clinical outcomes, and economics of the double-row repair technique reinforced with modified Mason–Allen and simple sutures [51].

Clinical Presentation

Postage stamp fractures are frequently reported in patients who are male, age 25 years or younger, and participants in sporting activities [10]. These fractures are also frequently reported in injuries initially stabilized with 3 or more anchors or conventional knot-tying anchors [10], as well as in fractures that experienced osteolysis around anchor sites [10]. Routine radiographic follow-up is recommended after the use of metallic anchors to ensure the identification of early failure by anchor pullout [11]. Overtightening sutures may cause anchor pull-out [13].

Long-term Outcomes and Revision Patterns

At a mean follow-up of 18.3 years, arthroscopic single-row suture anchor fixation of bony Bankart lesions demonstrated low rates of recurrent dislocation and subjective instability [18]. Good longterm results were observed after arthroscopic Bankart repair in patients older than 20 years with 3 or more suture anchors used [21]. In 22 revision rotator cuff repair cases, modes of failure included tendon pulling through suture, tear in a new location, suture anchors pulling out of bone, and suture breakage or suture knot [23]. Specifically, in 2 revision cases, the rotator cuff tendon tear was identified in a new location adjacent to the area that was previously repaired [23], and in 1 of 22 revision rotator cuff repairs, a loose suture anchor was associated with the failure [23].

Complications and Device-Specific Issues

Disadvantages of a 4.0-mm-size knotless all-suture anchor include a possible large defect if the anchor were to pull out [26], and no clear differentiator between the looped and non-looped end of the passing suture in the joint [26]. Osteolyses, sometimes major, have been highlighted following surgeries using resorbable anchors [24]. In a study of 52 patients who underwent shoulder stabilization with PLLA tacks, a 19% complication rate was noted [47]. All 10 patients with complications presented within 8 months of the index surgical procedure with complaints of pain and stiffness [47]. On repeat arthroscopy, all 10 patients with complications had synovitis [47], 9 patients had implant debris [47], and 6 patients had evidence of full-thickness chondral injury [47]. In a series of 6 patients with failed bioabsorbable tacks made of PLLA used for superior labral repair, all patients had MRI identification of the broken or dislodged tacks [47]. All patients in this series required repeat arthroscopic reevaluation for fragment removal, evaluation of the biceps anchor, and restabilization of the superior labrum [47]. No inflammatory reactions were documented in the studied biodegradable lactide-containing suture anchors [50].

Investigations

Plain radiography: Routine radiographic follow-up is recommended after the use of metallic suture anchors to identify early failure by anchor pullout [11]. The standard shoulder series should include a true AP view in the scapular plane, an AP view, an axillary view, and a scapular Y view [90]. The axillary view is necessary for the evaluation of glenohumeral joint instability and enables determination of the humeral head position in the glenoid fossa [90]. Secondary degenerative joint disease with a central humeral defect from contact with suture anchors can be observed on anteroposterior and axillary radiographs [40]. Various bone changes are seen on radiographs after implantation of an all-suture anchor, but these changes are not clinically relevant [1]. Radiolucent rings around bioabsorbable anchors after rotator cuff repair are not associated with clinical outcomes [129]. Most radiolucent rings around bioabsorbable anchors appeared at anteromedial anchors, indicating that mechanical factors may play a role for the radiolucencies [129]. Osteolysis, sometimes major, has been reported following surgeries using resorbable anchors [24].

CT: CT imaging is frequently used to evaluate fractures of the shoulder, assess for bony lesions in recurrent instability cases, or for preoperative templating for shoulder arthritis [89]. CT with three-dimensional reconstructions is the advanced imaging study of choice for determining the extent of glenoid bone loss in the setting of shoulder instability [90].

MRI: MRI is the modality of choice for evaluating the rotator cuff, biceps, and subacromial bursa [89]. All-suture anchors do not distort MRI, allowing the surgeon to better assess the quality of repair [42]. All-suture anchor fixation facilitates postoperative imaging on magnetic resonance imaging [48]. T1-weighted MRI can reveal Hill-Sachs lesions and is often used with magnetic resonance arthrograms to provide a more detailed picture of the joint surfaces [89]. T2-weighted MRI provides better visualization of full thickness rotator cuff tears [89]. MR arthrography is considered the benchmark for evaluation for labral tears and rarely is indicated for evaluation of rotator cuff pathology [89]. Magnetic resonance accuracy in identifying labral and rotator cuff tears in the literature ranges from 70% to 100% [87]. In a meta-analysis of 6 studies including 4,667 shoulders, MRA had greater diagnostic test accuracy for the detection of glenoid labral lesions than MRI, with MRA sensitivity of 88% and specificity of 93% versus MRI sensitivity of 76% and specificity of 87% [87]. MRI data suggest improved healing rates in single-row repairs in the entire patient population [145]. Double-row repair showed improved radiographic healing when similar-sized tears were compared [145]. Radiographic healing is significantly improved in double-row repairs [148].

Ultrasonography: Ultrasonography is a low-cost alternative to MRI and arthrography for evaluating both skeletal and soft-tissue structures of the shoulder [89]. Ultrasonography can provide immediate, real-time visualization of the rotator cuff, biceps tendon, and calcific deposits [89]. Ultrasonography is a simple and accurate test for identifying rotator cuff tears and calcific tendinitis [80]. The sensitivity of ultrasonography for the detection of full-thickness rotator cuff tears is 98% [90]. The specificity of ultrasonography for the detection of full-thickness rotator cuff tears is 80% [90]. The positive predictive value of ultrasonography for the detection of full-thickness rotator cuff tears is 90% [90]. The negative predictive value of ultrasonography for the detection of full-thickness rotator cuff tears is 95% [90]. The accuracy of ultrasonography for the detection of full-thickness rotator cuff tears is 94% [90]. Ultrasonography is highly operator dependent and is not as useful for evaluating labral tears or rotator cuff tears that are very small or larger than 3 cm [89].

Treatment

Operative

Indications: Arthroscopic suture anchor-based repair is indicated for posterior labral pathology, demonstrating lower recurrence rates and increased patient-reported outcomes compared to suture fixation alone or open surgical interventions [43]. For open Bankart repair, suture anchors are a standard fixation method, though recurrence rates vary across studies, ranging from 5.9% to 10% [45]. In contrast, the conventional transosseous suture technique for open Bankart repair carries a lower recurrence rate of 2.2% [45].

Surgical Approach / Technique: The choice between knotless and knotted systems does not significantly alter clinical outcomes or return-to-sport rates following arthroscopic anterior labral repair [29]. Knotless and knotted all-suture anchors yield comparable results in this setting [29]. In open Bankart operations, suture anchors simplify the procedure without introducing specific complications associated with modified techniques [12]. For intraoperative anchor pullout, bar anchoring with a threaded Steinmann pin is a viable solution that utilizes primary anchor sites with acceptable results [22]. The modified arthroscopic en masse suture bridge technique is effective for repairing delaminated rotator cuff tears, yielding favourable clinical outcomes comparable to those of nondelaminated tears [107].

Implant Selection: All-suture anchors possess similar or better mechanical properties than regular anchors, with a low-profile design that preserves bone tissue [5]. Biomechanical studies indicate similar properties between knotless and knotted anchors, as well as between all-suture and hard-bodied anchors in Bankart repair models [29]. Prior work demonstrates similar load-to-failure outcomes for “all-suture” suture anchor fixation compared to traditional suture anchors [43]. All-suture anchors have a smaller diameter than traditional hard-bodied anchors, allowing for more points of fixation over a given surface area and improved preservation of glenoid bone stock [29]. Advantages of knotless suture anchors include decreased surgical time, ease of use, improved ability to tension the repair, and low-profile fixation that alleviates concerns for articular cartilage abrasion from intraarticular knots [29]. The Knotless FiberTak is equipped with a shuttling suture mechanism, which allows the repair suture to be passed through its locking mechanism [53]. Bone fixation seems to be improved with the knotless anchor systems in arthroscopically assisted latissimus dorsi transfer fixation [3].

Adjuncts: Double-row suture anchor fixation does not improve clinical outcome according to published literature, although some studies report it may improve tendon healing [49]. Tape-like double-row suture bridge repair was clinically non-superior to conventional double-row suture bridge repair despite theoretical biomechanical advantages [61]. However, double-row suture bridge rotator cuff repair with all-suture anchors for medial row fixation demonstrated excellent short-term clinical outcomes and substantial improvements [38]. Arthroscopic tendon-to-bone transfer (ATSCR) outcomes are not inferior to anchor-based techniques and involve less suture tangling, more precise suture pass points, and lower costs [8]. This system facilitates robust and reproducible repairs while substantially reducing procedural costs compared to suture anchor techniques [37].

Complications and Risks: Complications related to suture anchors and tacks are categorized as technique-related or device-related issues, with prevention dependent on the surgeon's familiarity with the devices and knowledge of their indications and limitations [9]. Postage stamp fractures were reported frequently in patients who were male, age 25 years or younger, and participants in sporting activities, and in fractures initially stabilized with 3 or more anchors or conventional knot-tying anchors or that experienced osteolysis around anchor sites [10]. Bioabsorbable PLDLA anchor material led to significantly more SLAP repair failures and reoperations compared with nonabsorbable suture anchors [143]. Disadvantages of the 4.0-mm-size knotless all-suture anchor include a possible large defect if the anchor were to pull out and no clear differentiator between the looped and non-looped end of the passing suture in the joint [26]. In a series of 72 patients treated with open repair using Mitek anchors and capsular shift, there were 2 reluxations after 9 and 12 months and one instance of anchor loosening seen on X-ray without recurrent instability [81].

Other Considerations: Careful attention to proper anchor insertion techniques can limit the potential for complications, and newer materials may address concerns of biocompatibility and material strength [6]. A new resorbable anchor offers the possibility of greater biocompatibility than currently available absorbable suture anchors [55]. This method can be applied in any similar case, regardless of suture anchor type [19]. It requires only 2 suture anchors, fewer than for double-row repair, and allows slow, controlled compression of the fracture during reduction [7]. Radiological review of patients who underwent “allsuture” suture anchor fixation demonstrated similar healing rates to traditional suture anchor counterparts without bony reactions or cystic changes [43]. Routine radiographic follow-up is recommended after use of metallic anchors to ensure identification of early failure by anchor pullout [11].

Complications

General Classification and Prevention

Complications associated with suture anchors and tacks are categorized as either technique-related or device-related issues [9]. Suture anchor designs and materials have evolved specifically to minimize these complications [140].

Complications from metal staples, including dislocation of these staples, occur in up to 12% of cases [133].

Osteolyses, sometimes major, have been reported following surgeries using resorbable anchors [24]. In a series of 52 patients undergoing shoulder stabilization with PLLA tacks, a 19% complication rate (10/52) was noted [47]. All 10 patients presenting with complications did so within 8 months of the index surgical procedure [47]. On repeat arthroscopy, all 10 patients had synovitis [47], 9 of the 10 had implant debris [47], and 6 of the 10 had evidence of full-thickness chondral injury [47]. Three patients with chondral damage did not experience significant pain relief after debridement and lavage [47]. Additionally, six patients with failed bioabsorbable tacks made of PLLA used for superior labral repair required repeat arthroscopic reevaluation for fragment removal [47]. No inflammatory reactions were documented in studied biodegradable lactide-containing suture anchors [50]. Radiological review of patients who underwent “all-suture” suture anchor fixation demonstrated similar healing rates to traditional suture anchor counterparts without bony reactions or cystic changes [43].

“Postage stamp” fractures were reported frequently in fractures initially stabilized with 3 or more anchors [10] and in fractures initially stabilized with conventional knot-tying anchors [10]. Complications such as needle breakage and fractures of the glenoid rim while drilling the holes have also been reported [133].

Suture anchor complications, such as prolapse, are a concern in medical practice [118].

Tying arthroscopic knots for fixation is a common site of fixation failure [29]. Intraarticular knots raise concerns for articular cartilage abrasion [29]. ATSCR involves less suture tangling compared to anchor-based techniques [8].

Recurrence and Instability

The recurrence rate for open Bankart repair using the Statak device was 8.7% (2 resubluxations out of 23 operations) [45]. The recurrence rate for open Bankart repair using the Mitek Gl anchor was 5.9% (1 recurrence out of 17 cases) with 12 to 24 months of follow-up [45], 7.1% (2 of 28 cases) with a follow-up of 33 to 61 months [45], and 10% (2 redislocations in 20 patients) [45]. The failure rate for open Bankart repair using Statak suture anchors was 20% [133]. The failure rate for open Bankart repair using Mitek GI anchors was 7% (2 of 28 patients) with a follow-up of 42 months [133].

Revision Surgery Outcomes

Of the original 342 patients in a rotator cuff repair study, 21 (6%) subsequently underwent a revision rotator cuff repair by the same surgeon [23]. In a study of 22 revision rotator cuff repair cases, modes of failure included tendon pulling through suture, tear in a new location, suture anchors pulling out of bone, and suture breakage or suture knot [23]. In two cases of revision rotator cuff repair, the rotator cuff tendon tear was identified in a new location adjacent to the area that was previously repaired [23].

Recovery

Other Considerations: The recovery profile for suture anchor repairs is characterized by specific material-related changes and long-term surveillance requirements rather than a standardized, evidence-based timeline for activity restriction. Resorbable anchors are associated with osteolyses, which can be major in extent [24]. Newer materials are being developed to address concerns regarding biocompatibility and material strength for these bioabsorbable devices [6].

Anchor integrity is a critical component of the recovery phase. The security of a 1.8-mm all-suture anchor at the time of implantation is exceptionally reliable, with no demonstrable learning curve in the hands of an experienced surgeon [151]. Intraoperatively, secure fixation of soft tissue to bone was accomplished in every case, and follow-up x-ray films showed that 65 of the implants remained securely anchored in bone [27].

Long-term clinical outcomes indicate that following arthroscopic double-row suture-bridge rotator cuff repair, longer follow-up time was associated with better patient-reported outcomes in both anchor type groups [56]. However, long-term follow-up will be necessary to determine if the durability of these repairs and the structural integrity of these constructs maintain their performance over time [109]. Similarly, long term followup is necessary to determine the effectiveness of capsulorrhaphy using non absorbable suture anchors [150].

Complications and functional changes are monitored during recovery. Despite the increased incidence of a postoperative complication after knotted anchor fixation compared with knotless anchor fixation, multiple regression analysis showed that anchor design and anchor position were not significantly predictive of the incidence of complications [57]. In cases of metal-free autograft bone block suture tape cerclage for recurrent anterior shoulder instability, external rotation at the side and in 90º of abduction decreased 4º and 3º respectively [59].

Key Evidence

  • [L3] Various bone changes are seen after implantation of an all-suture anchor, but these are not clinically relevant. [1] (10.1177/17585732221127433)
  • [L5] The goals of this article are to review the biological and biomechanical factors that influence the successful evaluation and use of suture anchors and tacks. [2] (10.1177/0363546505282621)
  • [Paper] Bone fixation seems to be improved with the knotless anchor systems. [3] (10.1002/jor.24963)
  • [L5] The small, 2.6-mm all-suture anchor mitigates risks and difficulties associated with other fixation devices while providing ease and efficiency for the surgeon. [4] (10.1016/j.eats.2023.02.030)
  • [L4] All-suture anchors have similar or better mechanical properties than regular anchors, with low-profile design preserving bone tissue. [5] (10.1016/j.asmr.2020.02.007)
  • [L4] Careful attention to proper anchor insertion techniques can limit the potential for complications, and newer materials may address concerns of biocompatibility and material strength. [6] (10.1016/j.arthro.2008.08.018)
  • [L5] It requires only 2 suture anchors, fewer than for double-row repair, and allows slow, controlled compression of the fracture during reduction. [7] (10.1016/j.eats.2024.103297)
  • [L5] The outcomes of ATSCR are not inferior to those of anchor-based techniques and involve less suture tangling, more precise suture pass points, and lower costs. [8] (10.1016/j.eats.2023.03.022)
  • [L4] Complications related to suture anchors and tacks can be categorized as technique-related or device-related issues, and prevention depends on the surgeon's familiarity with the devices and knowledge of their indications and limitations. [9] (10.1177/0363546505284240)
  • [L4] Postage stamp fractures were reported frequently in patients who were male, age 25 years or younger, and participants in sporting activities and in fractures initially stabilized with 3 or more anchors or conventional knot-tying anchors or that experienced osteolysis around anchor sites. [10] (10.1016/j.arthro.2019.02.047)
  • [L3] We recommend routine radiographic follow-up after use of metallic anchors to ensure identification of early failure by anchor pullout. [11] (10.1016/j.arthro.2009.08.015)
  • [L3] The suture anchors were found to simplify the procedure, and no specific complications as a result of the modified technique were seen. [12] (10.1016/s1058-2746(96)80444-7)
  • [L5] The authors summarize the pearls and pitfalls of the new technique, noting that overtightening sutures may cause anchor pull-out. [13] (10.1016/j.eats.2023.11.019)
  • [L5] This technique avoids a difficult trajectory and large-diameter bone punches during anchor insertion seen with other techniques, providing a reproducible and secure fixation. [14] (10.1016/j.eats.2024.103354)
  • [L5] The bearing area of a suture anchor can be used to predict the pullout strength independent of design in a soft-bone model. [15] (10.1002/jor.20856)
  • [L3] The repair technique (knotted suture-bridging or knotless tape-bridging) did not have a significant effect on the final outcome at a mean of 2.9 years. [16] (10.1007/s00402-017-2750-7)
  • [L5] The TOAK technique is a biomechanically strong and secure cuff fixation that eliminates the potential risk for loose hardware in the subacromial space. [17] (10.1097/00132589-200412000-00004)
  • [L4] At a mean follow-up of 18.3 years, arthroscopic single-row suture anchor fixation of bony Bankart lesions demonstrated low rates of recurrent dislocation and subjective instability. [18] (10.1002/arj.70565)
  • [L5] This method can be applied in any similar case, regardless of suture anchor type. [19] (10.1142/s242483551872030x)
  • [L2] The Micro bone suture anchor provides a stronger tendon to bone repair than the pullout button or the Mini anchor. [20] (10.1053/jhsu.2001.24135)
  • [L4] Good longterm results were observed after arthroscopic Bankart repair in patients older than 20 years with 3 or more suture anchors used. [21] (10.1016/j.jse.2018.09.027)
  • [L3] While no single technique is overwhelmingly superior, bar anchoring with a threaded Steinmann pin is a viable solution that utilizes primary anchor sites with acceptable results. [22] (10.1007/s00167-020-05935-4)
  • [L3] [23] (10.1067/mse.2003.21)
  • [L4] The study highlights osteolyses, sometimes major, following surgeries using resorbable anchors. [24] (10.1016/j.main.2013.10.109)
  • [L5] Transosseous suture fixation is not necessarily a "weak link" of repairs. [25] (10.1016/s1058-2746(96)80169-8)
  • [L5] However, disadvantages include a possible large defect if the 4.0-mm-size anchor were to pull out and no clear differentiator between the looped and non-looped end of the passing suture in the joint. [26] (10.1016/j.eats.2025.103598)
  • [L4] Intraoperatively, secure fixation of soft tissue to bone was accomplished in every case, and follow-up x-ray films showed that 65 of the implants remained securely anchored in bone. [27] (10.1016/0363-5023(94)90201-1)
  • [L1] [28] (10.1016/j.arthro.2013.03.076)
  • [L3] [29] (10.1177/17585732261472078)
  • [L4] The shoulder function improved in all patients after the loosened anchor or screw was removed, but the cartilage damage might progress and cause symptoms in the future. [31] (10.1016/s1058-2746(98)90208-7)
  • [L5] Deep placement of suture anchors for increased purchase caused greater displacement and is not recommended. [32] (10.1016/j.jse.2005.03.011)
  • [L3] The results suggest that suture anchor fixation would be stronger than screw fixation immediately after surgery for humeral GT fracture. [33] (10.1002/jor.21530)
  • [L5] There has been a major shift from metallic to bioabsorbable anchors due to higher complication rates with metallic anchors and lower rates with bioabsorbable ones. [35] (10.1016/j.arthro.2007.05.011)
  • [L5] This method is particularly useful in large tears with multiple anchors or when suture limbs are entangled. [36] (10.1016/j.eats.2022.12.007)
  • [L5] This system facilitates robust and reproducible repairs while substantially reducing procedural costs compared to suture anchor techniques. [37] (10.1016/j.eats.2025.103566)
  • [L4] Our study demonstrated excellent short-term clinical outcomes and substantial improvements for patients undergoing double-row suture bridge RCR with all-suture anchors for medial row fixation. [38] (10.1177/23259671231192134)
  • [L3] [42] (10.1177/2325967126s00050)
  • [L5] [43] (10.1016/j.eats.2023.03.011)
  • [Paper] This technique is an advance in rotator cuff surgery and will help to facilitate the achievement of the main goals of rotator cuff surgery: pain reduction and restoration of shoulder functionality. [44] (10.1097/bte.0000000000000064)
  • [L3] [45] (10.1016/s1058-2746(99)90052-6)
  • [L5] [47] (10.1177/0363546511417573)
  • [L5] All-suture anchor fixation preserves glenoid bone stock, facilitates postoperative imaging on magnetic resonance imaging, and offers a biomechanically stable fixation. [48] (10.1016/j.eats.2025.103636)
  • [L2] At present, the data in the published literature do not support the use of DR suture anchor fixation to improve clinical outcome, but there are some studies that report that DR suture anchor fixation may improve tendon healing. [49] (10.1016/j.arthro.2009.02.005)
  • [L4] No inflammatory reactions were documented in the studied anchors. [50] (10.1016/j.arthro.2014.02.011)
  • [L5] Future studies are required to validate the biomechanical principles, clinical outcomes, and economics of this technique. [51] (10.1016/j.eats.2022.08.039)
  • [L4] [52] (10.1002/atn2.70169)
  • [L5] [53] (10.1016/j.eats.2024.102909)
  • [L5] This new anchor offers the possibility of greater biocompatibility than currently available absorbable suture anchors. [55] (10.1067/mse.2001.113085)
  • [L2] Following arthroscopic double-row suture-bridge RCR, longer follow-up time was associated with better patient-reported outcomes (PROs) in both anchor type groups. [56] (10.1016/j.xrrt.2025.100639)
  • [L4] Despite the increased incidence of a postoperative complication after knotted anchor fixation compared with knotless anchor fixation, multiple regression analysis showed that anchor design and anchor position were not significantly predictive of the incidence of complications. [57] (10.1177/2325967120911361)
  • [L3] External rotation at the side and in 90º of abduction decreased 4º and 3º respectively. [59] (10.1016/j.jseint.2024.08.159)
  • [L5] This arthroscopic variation of a well-known technique is simple and readily applied in any arthroscopic shoulder surgery practice. [60] (10.1016/j.eats.2023.02.024)
  • [L3] Tape-like double-row suture bridge repair was clinically non-superior to conventional double-row suture bridge repair despite theoretical biomechanical advantages. [61] (10.1007/s00167-023-07454-4)
  • [L3] [81] (10.1016/s1058-2746(96)80394-6)
  • [L5] This all-suture tape cerclage technique for the fixation of high-grade AC joint separations was designed to capitalize on the physical properties of this recently developed suture material while minimizing the risk of complications inherent to currently popular procedures. [102] (10.1016/j.eats.2024.103184)
  • [L2] The current literature reveals that the biomechanical properties of a double-row rotator cuff repair are superior to a single-row repair. [104] (10.1016/j.jse.2009.07.002)
  • [L3] The stump classification was significantly correlated with retear in the suture-bridge and double-row repair techniques. [105] (10.1007/s00167-020-06415-5)
  • [L5] [106] (10.1016/j.arthro.2010.02.007)
  • [L4] The modified arthroscopic en masse suture bridge technique was effective for repairing delaminated RCTs, yielding favourable clinical outcomes comparable to those of nondelaminated tears. [107] (10.1002/ksa.12412)
  • [L5] The technique described simplifies graft delivery and suture management by using a graft inserter and a self-punching, knotless technique of graft fixation. [108] (10.1016/j.eats.2022.08.060)
  • [L4] Long-term follow-up will be necessary to determine if the durability of these repairs and the structural integrity of these constructs maintain their performance over time. [109] (10.1177/0363546510397725)
  • [L4] By addressing the dual imperatives of tissue preservation and biomechanical stability, this technique expands the surgical armamentarium for managing complex subscapularis pathology while mitigating risks associated with conventional repair strategies. [111] (10.1016/j.eats.2025.103736)
  • [L5] [112] (10.1016/j.eats.2025.103762)
  • [L4] Although double-row fixation restores the rotator cuff footprint and shows biomechanical superiority, there is no evidence that it results in better functional outcomes compared to single-row fixation. [113] (10.1016/j.arthro.2007.01.031)
  • [L5] The approach is reproducible, minimally invasive, and suitable for degenerative or inflamed tendons, providing biomechanical stability and improved functional outcomes. [114] (10.1002/atn2.70092)
  • [L3] The augmentation also resulted in improved shoulder function and greater range of motion compared with the nonaugmented group. [116] (10.1016/j.jse.2023.08.002)
  • [L4] Current level II to IV clinical and biomechanical evidence supports knotless arthroscopic Bankart repair as an effective procedure with excellent functional outcomes and low rates of complication and reoperation. [117] (10.1016/j.asmr.2021.01.014)
  • [Case_report] Suture anchor complications, such as prolapse, are a concern in medical practice. [118] (10.1186/s12891-024-07476-0)
  • [L3] At 24 weeks, forward flexion and shoulder strength in abduction, adduction, external rotation, internal rotation, and lift-off improved compared with preoperative levels. [120] (10.1097/bte.0000000000000186)
  • [L1] On the basis of the included cadaveric studies, rotator cuff tendons repaired via the knotted TOE technique display superior time-zero biomechanical properties, including greater ultimate load to failure, compared with rotator cuffs repaired via the knotless TOE technique. [121] (10.1016/j.arthro.2021.09.017)
  • [L3] [122] (10.1016/j.arthro.2019.03.032)
  • [L4] Significant improvements were observed across pain, range of motion, strength, and functional outcomes. [125] (10.1016/j.jseint.2026.101778)
  • [L5] The knotless, low-profile design, using a readily available biomechanically tested anchor, eliminates a prominent knot stack superiorly and decreases the risk of stress riser in the acromion and distal clavicle compared to larger, solid body anchors. [126] (10.1016/j.eats.2024.103226)
  • [L5] This technique combines the biomechanical strength of using multiple smaller anchors with knotless fixation. [127] (10.1016/j.eats.2023.03.016)
  • [L5] The construct aims to maximize footprint coverage, increase tendon thickness and effective length, and restore superior stability of the glenohumeral joint while relying on widely available implants and reproducible steps. [128] (10.1002/atn2.70153)
  • [Paper] Most radiolucent rings appeared at anteromedial anchors, indicating that mechanical factors may play a role for the radiolucencies. [129] (10.1007/s00402-017-2772-1)
  • [L1] However, there was conflicting evidence supporting knotless or knotted anchors in the biomechanical studies, though operative times may be reduced with the use of knotless anchors. [130] (10.1016/j.arthro.2020.11.056)
  • [L3] All-suture anchors with various deployment configurations produced different tunnel enlargements, with the diameter more pronounced at the inferior region of the anterior glenoid. [131] (10.1177/23259671251319533)
  • [L2] [133] (10.1016/s1058-2746(03)00032-6)
  • [L5] Due to technical efficiency of the proposed procedure, and ability to still maintain a strong biomechanical construct, we believe that a single anchor medial-row construct to be a promising technique in the repair of even massive rotator cuff tendon tears. [136] (10.1016/j.eats.2022.03.035)
  • [L5] The double-barrel knot is a new arthroscopic sliding knot for soft tissue reconstruction and repair procedures in arthroscopic and open shoulder surgery. [138] (10.1097/bth.0b013e318285b4e4)
  • [L4] Suture anchors have revolutionized upper limb surgery by enabling arthroscopic soft tissue-to-bone repair, with designs and materials evolving to minimize complications. [140] (10.1016/j.hcl.2012.08.002)
  • [L3] Bioabsorbable PLDLA anchor material led to significantly more SLAP repair failures and reoperations compared with nonabsorbable suture anchors. [143] (10.1016/j.arthro.2011.06.021)
  • [L3] Although MRI data suggest improved healing rates in SR repairs in the entire patient population, DR repair showed improved radiographic healing when similar-sized tears were compared. [145] (10.1016/j.arthro.2010.03.013)
  • [L3] However, radiographic healing is significantly improved in the double row repairs. [148] (10.1016/j.arthro.2009.04.005)
  • [L4] Although the early results of this study are promising, long term followup is necessary to determine the effectiveness of this procedure. [150] (10.1016/s1058-2746(96)80119-4)
  • [L4] The data support that the security of this particular 1.8-mm all-suture anchor at the time of implantation is exceptionally reliable, and there is no demonstrable learning curve in employing the product in the hands of an experienced surgeon. [151] (10.1016/j.arthro.2017.09.049)

See Also

References

[1] Clinical and radiological outcome of all-suture anchors in shoulder and elbow surgery. Shoulder & Elbow. 2022. DOI: 10.1177/17585732221127433

[2] Suture Anchors and Tacks for Shoulder Surgery, Part 1. The American Journal of Sports Medicine. 2005. DOI: 10.1177/0363546505282621

[3] Biomechanical analysis of arthroscopically assisted latissimus dorsi transfer fixation for irreparable posterosuperior rotator cuff tears—Knotless versus knotted anchors. Journal of Orthopaedic Research. 2020. DOI: 10.1002/jor.24963

[4] Subpectoral Biceps Tenodesis Using an All‐Suture Knotless Anchor. Arthroscopy Techniques. 2023. DOI: 10.1016/j.eats.2023.02.030

[5] The Clinical and Biomechanical Performance of All‐Suture Anchors: A Systematic Review. Arthroscopy, Sports Medicine, and Rehabilitation. 2020. DOI: 10.1016/j.asmr.2020.02.007

[6] Bioabsorbable Anchors in Glenohumeral Shoulder Surgery. Arthroscopy. 2009. DOI: 10.1016/j.arthro.2008.08.018

[7] Arthroscopic Bony Bankart Lesion Fixation Using Knotless Double‐Loaded Suture Anchors and Buttress Sutures. Arthroscopy Techniques. 2024. DOI: 10.1016/j.eats.2024.103297

[8] Arthroscopic Transosseous Superior Capsular Reconstruction. Arthroscopy Techniques. 2023. DOI: 10.1016/j.eats.2023.03.022

[9] Suture Anchors and Tacks for Shoulder Surgery, Part II. The American Journal of Sports Medicine. 2006. DOI: 10.1177/0363546505284240

[10] “Postage Stamp” Fractures: A Systematic Review of Patient and Suture Anchor Profiles Causing Anterior Glenoid Rim Fractures After Bankart Repair. Arthroscopy. 2019. DOI: 10.1016/j.arthro.2019.02.047

[11] The Incidence of Early Metallic Suture Anchor Pullout After Arthroscopic Rotator Cuff Repair. Arthroscopy. 2010. DOI: 10.1016/j.arthro.2009.08.015

[12] The use of suture anchors in open bankart operation. Journal of Shoulder and Elbow Surgery. 1996. DOI: 10.1016/s1058-2746(96)80444-7

[13] A Knotless Single‐Row Suture Bridge Technique to Repair a Partial Articular Supraspinatus Tendon Avulsion‐Type Rotator Cuff Tear: Surgical Technique. Arthroscopy Techniques. 2024. DOI: 10.1016/j.eats.2023.11.019

[14] Arthroscopic Retro All‐Suture Anchor Technique for Subscapularis Repair Using a Posterior Cruciate Ligament Femoral Guide. Arthroscopy Techniques. 2024. DOI: 10.1016/j.eats.2024.103354

[15] Bearing area: A new indication for suture anchor pullout strength?. Journal of Orthopaedic Research. 2009. DOI: 10.1002/jor.20856

[16] Predictors of outcomes after arthroscopic transosseous equivalent rotator cuff repair in 155 cases: a propensity score weighted analysis of knotted and knotless self-reinforcing repair techniques at a minimum of 2 years. Archives of Orthopaedic and Trauma Surgery. 2017. DOI: 10.1007/s00402-017-2750-7

[17] Transosseous Anchor Double Knot (TOAK) Technique for Rotator Cuff Repair. Techniques in Shoulder and Elbow Surgery. 2004. DOI: 10.1097/00132589-200412000-00004

[18] Long‐Term Clinical Outcomes of Arthroscopic Suture Anchor Fixation of Bony Bankart Lesions at a Mean 18‐Year Follow‐Up. Arthroscopy. 2026. DOI: 10.1002/arj.70565

[19] Repair Technique Using a Combination of Suture Anchor and Miniscrew in Flexor Digitorum Profundus Bony Avulsion Fracture with Bone Fragment in Zone 1. The Journal of Hand Surgery (Asian-Pacific Volume). 2018. DOI: 10.1142/s242483551872030x

[20] Bone suture anchors versus the pullout button for repair of distal profundus tendon injuries: A comparison of strength in human cadaveric hands. The Journal of Hand Surgery. 2001. DOI: 10.1053/jhsu.2001.24135

[21] Long-term clinical outcome of arthroscopic Bankart repair with suture anchors. Journal of Shoulder and Elbow Surgery. 2019. DOI: 10.1016/j.jse.2018.09.027

[22] Novel and reproducible technique coping with intraoperative anchor pullout during arthroscopic rotator cuff repair. Knee Surgery, Sports Traumatology, Arthroscopy. 2020. DOI: 10.1007/s00167-020-05935-4

[23] Mode of failure for rotator cuff repair with suture anchors identified at revision surgery. Journal of Shoulder and Elbow Surgery. 2003. DOI: 10.1067/mse.2003.21

[24] Utilisation des ancres résorbables en chirurgie de la main et ostéolyses secondaires. Chirurgie de la Main. 2013. DOI: 10.1016/j.main.2013.10.109

[25] Transosseous rotator cuff fixation: The weak link? A biomechanical evaluation. Journal of Shoulder and Elbow Surgery. 1996. DOI: 10.1016/s1058-2746(96)80169-8

[26] Arthroscopic Shoulder Labral Repair Using Knotless All‐Suture Anchor With Suture Tape. Arthroscopy Techniques. 2025. DOI: 10.1016/j.eats.2025.103598

[27] The mitek bone anchor: Application to the hand, wrist and elbow. The Journal of Hand Surgery. 1994. DOI: 10.1016/0363-5023(94)90201-1

[28] Outcomes of Single‐Row Versus Double‐Row Arthroscopic Rotator Cuff Repair: A Systematic Review and Meta‐Analysis of Current Evidence. Arthroscopy. 2013. DOI: 10.1016/j.arthro.2013.03.076

[29] Knotless and knotted all-suture anchors have comparable clinical outcomes and return to sport rates following arthroscopic anterior labral (Bankart) repair. Shoulder & Elbow. 2026. DOI: 10.1177/17585732261472078

[31] Cartilage injuries in the shoulder joint caused by migration of suture anchors or mini screw. Journal of Shoulder and Elbow Surgery. 1998. DOI: 10.1016/s1058-2746(98)90208-7

[32] Increasing the insertion depth of suture anchors for rotator cuff repair does not improve biomechanical stability. Journal of Shoulder and Elbow Surgery. 2005. DOI: 10.1016/j.jse.2005.03.011

[33] Suture anchor versus screw fixation for greater tuberosity fractures of the humerus—a biomechanical study. Journal of Orthopaedic Research. 2011. DOI: 10.1002/jor.21530

[35] The Use of Anchors in Shoulder Surgery: A Shift From Metallic to Bioabsorbable Anchors. Arthroscopy. 2007. DOI: 10.1016/j.arthro.2007.05.011

[36] A Safe Technique to Ensure Not to Unload a Suture Anchor: Suture Dyeing. Arthroscopy Techniques. 2023. DOI: 10.1016/j.eats.2022.12.007

[37] Arthroscopic Transosseous Rotator Cuff Repair Using Bone Tunneling Device. Arthroscopy Techniques. 2025. DOI: 10.1016/j.eats.2025.103566

[38] Outcomes of Double-Row Rotator Cuff Repair Using a Novel All-Suture Soft Anchor Medial Row. Orthopaedic Journal of Sports Medicine. 2023. DOI: 10.1177/23259671231192134

[40] Rockwood And Matsen S The Shoulder. Arthroscopic Management of Prearthritic and Arthritic Conditions of the Shoulder and the Postarthroplasty Shoulder > Radiographic Evaluation.

[42] Outcomes Following Usage of All Suture Anchors for Rotator Cuff Repair. Orthopaedic Journal of Sports Medicine. 2026. DOI: 10.1177/2325967126s00050

[43] Posterior Labral Repair Using Knotless “All‐Suture” Suture Anchors. Arthroscopy Techniques. 2023. DOI: 10.1016/j.eats.2023.03.011

[44] Double-Row, Transosseous-Equivalent Suture-Bridge Repair for Supraspinatus Tears: Power Up the Healing. Techniques in Shoulder & Elbow Surgery. 2016. DOI: 10.1097/bte.0000000000000064

[45] Recurrences after the open Bankart repair: A potential risk with use of suture anchors. Journal of Shoulder and Elbow Surgery. 1999. DOI: 10.1016/s1058-2746(99)90052-6

[47] Complications of Bioabsorbable Suture Anchors in the Shoulder. The American Journal of Sports Medicine. 2011. DOI: 10.1177/0363546511417573

[48] Shoulder Latarjet Procedure Using All‐Suture Anchor and Button Fixation. Arthroscopy Techniques. 2025. DOI: 10.1016/j.eats.2025.103636

[49] Does the Literature Support Double‐Row Suture Anchor Fixation for Arthroscopic Rotator Cuff Repair? A Systematic Review Comparing Double‐Row and Single‐Row Suture Anchor Configuration. Arthroscopy. 2009. DOI: 10.1016/j.arthro.2009.02.005

[50] Adverse Events Associated With Biodegradable Lactide‐Containing Suture Anchors. Arthroscopy. 2014. DOI: 10.1016/j.arthro.2014.02.011

[51] Arthroscopic Double‐Row Repair of Posterosuperior Rotator Cuff Tears: Suture Bridge Technique Reinforced With Modified Mason–Allen and Simple Sutures. Arthroscopy Techniques. 2022. DOI: 10.1016/j.eats.2022.08.039

[52] A “Suture Staple” Knotless Anchor Technique for Fixation of Tibial Tubercle Avulsion Fractures in Pediatric Patients. Arthroscopy Techniques. 2026. DOI: 10.1002/atn2.70169

[53] Knotless Tensionable Anchors: Versatility, Tips, and Tricks. Arthroscopy Techniques. 2024. DOI: 10.1016/j.eats.2024.102909

[55] A novel, resorbable suture anchor: Pullout strength from the human cadaver greater tuberosity. Journal of Shoulder and Elbow Surgery. 2001. DOI: 10.1067/mse.2001.113085

[56] Predictors of outcomes following double-row rotator cuff repair: an assessment of all-suture or solid medial row anchor utilization at a single high-volume institution. JSES Reviews, Reports, and Techniques. 2026. DOI: 10.1016/j.xrrt.2025.100639

[57] Complications and Return to Activity After Arthroscopic Repair of Isolated Type II SLAP Lesions: A Systematic Review Comparing Knotted Versus Knotless Suture Anchors. Orthopaedic Journal of Sports Medicine. 2020. DOI: 10.1177/2325967120911361

[59] Metal-Free Autograft Bone Block Suture Tape Cerclage For Recurrent Anterior Shoulder Instability: Outcomes, Graft Resorption Mapping And Return To Sport. JSES International. 2024. DOI: 10.1016/j.jseint.2024.08.159

[60] Transosseous Equivalent Technique for Bony Bankart Repair. Arthroscopy Techniques. 2023. DOI: 10.1016/j.eats.2023.02.024

[61] Rotator cuff repairs with all‐suture tape anchors: no difference in outcomes between with or without all‐suture tape anchors. Knee Surgery, Sports Traumatology, Arthroscopy. 2023. DOI: 10.1007/s00167-023-07454-4

[63] Rockwood And Matsen S The Shoulder. Shoulder and Elbow Specialty Clinic Workers’ Survey > ANATOMY.

[64] A Lange Medical Book Current Diagnosis Treatment In Orthopedics Fifth Edition. 2Musculoskeletal Trauma Surgery > SHOULDER AND ARM INJURIES.

[65] Rockwood And Green S Fractures In Adults. 29: Principles of Nonunion and Bone Defect Treatment > Applied Anatomy Related to Scapular Fractures.

[66] Aaos Comprehensive Orthopaedic Review 3. Anatomy of the Shoulder, Arm, and Elbow > I. Shoulder.

[67] Rockwood And Matsen S The Shoulder. Developmental Anatomy of the Shoulder and Anatomy of the Glenohumeral Joint > Bursae.

[70] Rockwood And Matsen S The Shoulder. Fractures, Dislocations, and Acquired Problems of the Shoulder in Children > FRACTURES OF THE PROXIMAL HUMERUS.

[71] Apley And Solomon S Concise System Of Orthopaedics And Trauma. DISORDERS OF THE ROTATOR CUFF.

[75] Campbell S Operative Orthopaedics 4 Volume Set. RECONSTRUCTIVE PROCEDURES OF THE SHOULDER AND ELBOW IN ADULTS > ANATOMY AND BIOMECHANICS.

[76] Campbell S Operative Orthopaedics 4 Volume Set. ANTERIOR CRUCIATE LIGAMENT RECONSTRUCTION WITH BONE-PATELLAR TENDON-BONE GRAFT > SHOULDER INJURIES > ANATOMY AND BIOMECHANICS.

[77] Miller S Review Of Orthopaedics. Genetics of musculoskeletal conditions and abnormalities are summarized in Table 1.27 > UPPER EXTREMITY > SHOULDER.

[80] Apley And Solomon S Concise System Of Orthopaedics And Trauma. INVESTIGATION.

[81] Bankart repair using a suture anchor: A comparison between an arthroscopic and open technique. Journal of Shoulder and Elbow Surgery. 1996. DOI: 10.1016/s1058-2746(96)80394-6

[87] Rockwood And Green S Fractures In Adults. 29: Principles of Nonunion and Bone Defect Treatment > Magnetic Resonance Imaging and Arthrography.

[89] Orthopaedic Knowledge Update 13 Ebook Without Multimedia. Shoulder Anatomy and Biomechanics, Clinical Evaluation, Imaging > Clinical Evaluation > Imaging.

[90] Aaos Comprehensive Orthopaedic Review 3. Imaging of the Shoulder and Elbow > I. Shoulder.

[102] Reconstruction of the Coracoclavicular Ligament Complex Utilizing an All‐Suture Tape Cerclage Technique. Arthroscopy Techniques. 2024. DOI: 10.1016/j.eats.2024.103184

[104] Double-row vs single-row rotator cuff repair: A review of the biomechanical evidence. Journal of Shoulder and Elbow Surgery. 2009. DOI: 10.1016/j.jse.2009.07.002

[105] Stump classification was correlated with retear in the suture‐bridge and double‐row repair techniques for arthroscopic rotator cuff repair. Knee Surgery, Sports Traumatology, Arthroscopy. 2021. DOI: 10.1007/s00167-020-06415-5

[106] Bridging Self‐Reinforcing Double‐Row Rotator Cuff Repair: We Really Are Doing Better. Arthroscopy. 2010. DOI: 10.1016/j.arthro.2010.02.007

[107] A modified arthroscopic en masse suture bridge repair is effective for delaminated rotator cuff tears. Knee Surgery, Sports Traumatology, Arthroscopy. 2024. DOI: 10.1002/ksa.12412

[108] Dermal Tuberoplasty for Irreparable Supraspinatus Tears Using Self‐Punching, Knotless Fixation. Arthroscopy Techniques. 2022. DOI: 10.1016/j.eats.2022.08.060

[109] Early Structural and Functional Outcomes for Arthroscopic Double-Row Transosseous-Equivalent Rotator Cuff Repair. The American Journal of Sports Medicine. 2011. DOI: 10.1177/0363546510397725

[111] Intra‐articular Subscapularis Repair Using Transtendon Lasso‐Loop Technique With All‐Suture Anchor. Arthroscopy Techniques. 2025. DOI: 10.1016/j.eats.2025.103736

[112] Zip‐tie Multidirectional Instability: A Technique for Progressive Tightening of Knotless Anchors in Multidirectional Instability. Arthroscopy Techniques. 2025. DOI: 10.1016/j.eats.2025.103762

[113] Clinical Evidence Shows No Difference Between Single‐ and Double‐Row Repair for Rotator Cuff Tears. Arthroscopy. 2007. DOI: 10.1016/j.arthro.2007.01.031

[114] Tensionable Knotless Tendon‐Sparing Biceps Rerouting Technique for Irreparable Rotator Cuff Tears. Arthroscopy Techniques. 2026. DOI: 10.1002/atn2.70092

[116] Outcomes of arthroscopic single-row repair alone vs. repair with human dermal allograft patch augmentation in patients with large to massive, posterosuperior rotator cuff tears: a retrospective comparative study. Journal of Shoulder and Elbow Surgery. 2024. DOI: 10.1016/j.jse.2023.08.002

[117] Excellent Functional Outcomes and Low Complication Rates Following Knotless Arthroscopic Bankart Repair: A Systematic Review of Clinical and Biomechanical Studies. Arthroscopy, Sports Medicine, and Rehabilitation. 2021. DOI: 10.1016/j.asmr.2021.01.014

[118] Extensor tendon rupture and preoperative mri confirmations of suture anchor prolapse: a case report and literature review. BMC Musculoskeletal Disorders. 2024. DOI: 10.1186/s12891-024-07476-0

[120] Short-Term to Mid-Term Outcomes of Arthroscopic Stabilization Using PEEK Knotless Anchors. Techniques in Shoulder & Elbow Surgery. 2020. DOI: 10.1097/bte.0000000000000186

[121] Knotted Transosseous‐Equivalent Technique for Rotator Cuff Repair Shows Superior Biomechanical Properties Compared With a Knotless Technique: A Systematic Review and Meta‐analysis. Arthroscopy. 2021. DOI: 10.1016/j.arthro.2021.09.017

[122] Perianchor Cyst Formation After Arthroscopic Rotator Cuff Repair Using All‐Suture–Type, Bioabsorbable‐Type, and PEEK‐Type Anchors. Arthroscopy. 2019. DOI: 10.1016/j.arthro.2019.03.032

[125] Undersurface rotator cuff repair for partial-thickness tears using the Opus knotless system: a prospective cohort of 1000 consecutive cases with a 1.4% retear rate. JSES International. 2026. DOI: 10.1016/j.jseint.2026.101778

[126] Acromioclavicular Joint Reconstruction With Acromioclavicular Ligament Augmentation Using a Knotless, All‐Suture Anchor Construct. Arthroscopy Techniques. 2024. DOI: 10.1016/j.eats.2024.103226

[127] Fixation of the Proximal Hamstring Tendon Using an All‐Suture Tensionable Knotless Technique. Arthroscopy Techniques. 2023. DOI: 10.1016/j.eats.2023.03.016

[128] Reinforcing Anterosuperior Medium‐to‐Large Rotator Cuff Repairs With the Long Head of the Biceps Tendon and Dermal Allograft Patch Using a Knotless Double‐Row Technique. Arthroscopy Techniques. 2026. DOI: 10.1002/atn2.70153

[129] Radiolucent rings around bioabsorbable anchors after rotator cuff repair are not associated with clinical outcomes. Archives of Orthopaedic and Trauma Surgery. 2017. DOI: 10.1007/s00402-017-2772-1

[130] Knotted Versus Knotless Anchors for Labral Repair in the Shoulder: A Systematic Review. Arthroscopy. 2020. DOI: 10.1016/j.arthro.2020.11.056

[131] All-Suture Anchor Deployment Configurations in Arthroscopic Bankart Repair: A Comparative Analysis of Clinical and Radiological Outcomes. Orthopaedic Journal of Sports Medicine. 2025. DOI: 10.1177/23259671251319533

[133] Results of suture anchor repair of anteroinferior shoulder instability: a prospective clinical study of 85 shoulders. Journal of Shoulder and Elbow Surgery. 2003. DOI: 10.1016/s1058-2746(03)00032-6

[136] Single Medial‐Row Anchor With Biceps Tenodesis in a Transosseous Double‐Row Construct for Massive Rotator Cuff Tear. Arthroscopy Techniques. 2022. DOI: 10.1016/j.eats.2022.03.035

[138] The “Double-Barrel” Knot. Techniques in Hand & Upper Extremity Surgery. 2013. DOI: 10.1097/bth.0b013e318285b4e4

[140] Use of Suture Anchors and New Suture Materials in the Upper Extremity. Hand Clinics. 2012. DOI: 10.1016/j.hcl.2012.08.002

[143] Poly‐L/D‐Lactic Acid Anchors Are Associated With Reoperation and Failure of SLAP Repairs. Arthroscopy. 2011. DOI: 10.1016/j.arthro.2011.06.021

[145] Comparative Analysis of Single‐Row Versus Double‐Row Repair of Rotator Cuff Tears. Arthroscopy. 2010. DOI: 10.1016/j.arthro.2010.03.013

[148] Functional Outcomes and Radiographic Appearance of Single versus Double Row Rotator Cuff Repair (SS‐04). Arthroscopy. 2009. DOI: 10.1016/j.arthro.2009.04.005

[150] Recurrent anterior dislocation of the shoulder: Results of capsulorrhaphy using non absorbable suture anchors. Journal of Shoulder and Elbow Surgery. 1996. DOI: 10.1016/s1058-2746(96)80119-4

[151] Acetabular All‐Suture Anchor for Labral Repair: Incidence of Intraoperative Failure due to Pullout. Arthroscopy. 2018. DOI: 10.1016/j.arthro.2017.09.049

Creative Commons BY-NC 4.0

CC Creative Commons licence
BY Attribution — you must credit the source
NC NonCommercial — not for commercial use

Attribution-NonCommercial 4.0 International


Creative Commons Corporation ("Creative Commons") is not a law firm and does not provide legal services or legal advice. Distribution of Creative Commons public licenses does not create a lawyer-client or other relationship. Creative Commons makes its licenses and related information available on an "as-is" basis. Creative Commons gives no warranties regarding its licenses, any material licensed under their terms and conditions, or any related information. Creative Commons disclaims all liability for damages resulting from their use to the fullest extent possible.

Using Creative Commons Public Licenses

Creative Commons public licenses provide a standard set of terms and conditions that creators and other rights holders may use to share original works of authorship and other material subject to copyright and certain other rights specified in the public license below. The following considerations are for informational purposes only, are not exhaustive, and do not form part of our licenses.

Considerations for licensors: Our public licenses are intended for use by those authorized to give the public permission to use material in ways otherwise restricted by copyright and certain other rights. Our licenses are irrevocable. Licensors should read and understand the terms and conditions of the license they choose before applying it. Licensors should also secure all rights necessary before applying our licenses so that the public can reuse the material as expected. Licensors should clearly mark any material not subject to the license. This includes other CC- licensed material, or material used under an exception or limitation to copyright. More considerations for licensors: wiki.creativecommons.org/Considerations_for_licensors

Considerations for the public: By using one of our public licenses, a licensor grants the public permission to use the licensed material under specified terms and conditions. If the licensor's permission is not necessary for any reason--for example, because of any applicable exception or limitation to copyright--then that use is not regulated by the license. Our licenses grant only permissions under copyright and certain other rights that a licensor has authority to grant. Use of the licensed material may still be restricted for other reasons, including because others have copyright or other rights in the material. A licensor may make special requests, such as asking that all changes be marked or described. Although not required by our licenses, you are encouraged to respect those requests where reasonable. More considerations for the public: wiki.creativecommons.org/Considerations_for_licensees


Creative Commons Attribution-NonCommercial 4.0 International Public License

By exercising the Licensed Rights (defined below), You accept and agree to be bound by the terms and conditions of this Creative Commons Attribution-NonCommercial 4.0 International Public License ("Public License"). To the extent this Public License may be interpreted as a contract, You are granted the Licensed Rights in consideration of Your acceptance of these terms and conditions, and the Licensor grants You such rights in consideration of benefits the Licensor receives from making the Licensed Material available under these terms and conditions.

Section 1 -- Definitions.

a. Adapted Material means material subject to Copyright and Similar Rights that is derived from or based upon the Licensed Material and in which the Licensed Material is translated, altered, arranged, transformed, or otherwise modified in a manner requiring permission under the Copyright and Similar Rights held by the Licensor. For purposes of this Public License, where the Licensed Material is a musical work, performance, or sound recording, Adapted Material is always produced where the Licensed Material is synched in timed relation with a moving image.

b. Adapter's License means the license You apply to Your Copyright and Similar Rights in Your contributions to Adapted Material in accordance with the terms and conditions of this Public License.

c. Copyright and Similar Rights means copyright and/or similar rights closely related to copyright including, without limitation, performance, broadcast, sound recording, and Sui Generis Database Rights, without regard to how the rights are labeled or categorized. For purposes of this Public License, the rights specified in Section 2(b)(1)-(2) are not Copyright and Similar Rights.

d. Effective Technological Measures means those measures that, in the absence of proper authority, may not be circumvented under laws fulfilling obligations under Article 11 of the WIPO Copyright Treaty adopted on December 20, 1996, and/or similar international agreements.

e. Exceptions and Limitations means fair use, fair dealing, and/or any other exception or limitation to Copyright and Similar Rights that applies to Your use of the Licensed Material.

f. Licensed Material means the artistic or literary work, database, or other material to which the Licensor applied this Public License.

g. Licensed Rights means the rights granted to You subject to the terms and conditions of this Public License, which are limited to all Copyright and Similar Rights that apply to Your use of the Licensed Material and that the Licensor has authority to license.

h. Licensor means the individual(s) or entity(ies) granting rights under this Public License.

i. NonCommercial means not primarily intended for or directed towards commercial advantage or monetary compensation. For purposes of this Public License, the exchange of the Licensed Material for other material subject to Copyright and Similar Rights by digital file-sharing or similar means is NonCommercial provided there is no payment of monetary compensation in connection with the exchange.

j. Share means to provide material to the public by any means or process that requires permission under the Licensed Rights, such as reproduction, public display, public performance, distribution, dissemination, communication, or importation, and to make material available to the public including in ways that members of the public may access the material from a place and at a time individually chosen by them.

k. Sui Generis Database Rights means rights other than copyright resulting from Directive 96/9/EC of the European Parliament and of the Council of 11 March 1996 on the legal protection of databases, as amended and/or succeeded, as well as other essentially equivalent rights anywhere in the world.

l. You means the individual or entity exercising the Licensed Rights under this Public License. Your has a corresponding meaning.

Section 2 -- Scope.

a. License grant.

1. Subject to the terms and conditions of this Public License, the Licensor hereby grants You a worldwide, royalty-free, non-sublicensable, non-exclusive, irrevocable license to exercise the Licensed Rights in the Licensed Material to:

a. reproduce and Share the Licensed Material, in whole or in part, for NonCommercial purposes only; and

b. produce, reproduce, and Share Adapted Material for NonCommercial purposes only.

2. Exceptions and Limitations. For the avoidance of doubt, where Exceptions and Limitations apply to Your use, this Public License does not apply, and You do not need to comply with its terms and conditions.

3. Term. The term of this Public License is specified in Section 6(a).

4. Media and formats; technical modifications allowed. The Licensor authorizes You to exercise the Licensed Rights in all media and formats whether now known or hereafter created, and to make technical modifications necessary to do so. The Licensor waives and/or agrees not to assert any right or authority to forbid You from making technical modifications necessary to exercise the Licensed Rights, including technical modifications necessary to circumvent Effective Technological Measures. For purposes of this Public License, simply making modifications authorized by this Section 2(a) (4) never produces Adapted Material.

5. Downstream recipients.

a. Offer from the Licensor -- Licensed Material. Every recipient of the Licensed Material automatically receives an offer from the Licensor to exercise the Licensed Rights under the terms and conditions of this Public License.

b. No downstream restrictions. You may not offer or impose any additional or different terms or conditions on, or apply any Effective Technological Measures to, the Licensed Material if doing so restricts exercise of the Licensed Rights by any recipient of the Licensed Material.

6. No endorsement. Nothing in this Public License constitutes or may be construed as permission to assert or imply that You are, or that Your use of the Licensed Material is, connected with, or sponsored, endorsed, or granted official status by, the Licensor or others designated to receive attribution as provided in Section 3(a)(1)(A)(i).

b. Other rights.

1. Moral rights, such as the right of integrity, are not licensed under this Public License, nor are publicity, privacy, and/or other similar personality rights; however, to the extent possible, the Licensor waives and/or agrees not to assert any such rights held by the Licensor to the limited extent necessary to allow You to exercise the Licensed Rights, but not otherwise.

2. Patent and trademark rights are not licensed under this Public License.

3. To the extent possible, the Licensor waives any right to collect royalties from You for the exercise of the Licensed Rights, whether directly or through a collecting society under any voluntary or waivable statutory or compulsory licensing scheme. In all other cases the Licensor expressly reserves any right to collect such royalties, including when the Licensed Material is used other than for NonCommercial purposes.

Section 3 -- License Conditions.

Your exercise of the Licensed Rights is expressly made subject to the following conditions.

a. Attribution.

1. If You Share the Licensed Material (including in modified form), You must:

a. retain the following if it is supplied by the Licensor with the Licensed Material:

i. identification of the creator(s) of the Licensed Material and any others designated to receive attribution, in any reasonable manner requested by the Licensor (including by pseudonym if designated);

ii. a copyright notice;

iii. a notice that refers to this Public License;

iv. a notice that refers to the disclaimer of warranties;

v. a URI or hyperlink to the Licensed Material to the extent reasonably practicable;

b. indicate if You modified the Licensed Material and retain an indication of any previous modifications; and

c. indicate the Licensed Material is licensed under this Public License, and include the text of, or the URI or hyperlink to, this Public License.

2. You may satisfy the conditions in Section 3(a)(1) in any reasonable manner based on the medium, means, and context in which You Share the Licensed Material. For example, it may be reasonable to satisfy the conditions by providing a URI or hyperlink to a resource that includes the required information.

3. If requested by the Licensor, You must remove any of the information required by Section 3(a)(1)(A) to the extent reasonably practicable.

4. If You Share Adapted Material You produce, the Adapter's License You apply must not prevent recipients of the Adapted Material from complying with this Public License.

Section 4 -- Sui Generis Database Rights.

Where the Licensed Rights include Sui Generis Database Rights that apply to Your use of the Licensed Material:

a. for the avoidance of doubt, Section 2(a)(1) grants You the right to extract, reuse, reproduce, and Share all or a substantial portion of the contents of the database for NonCommercial purposes only;

b. if You include all or a substantial portion of the database contents in a database in which You have Sui Generis Database Rights, then the database in which You have Sui Generis Database Rights (but not its individual contents) is Adapted Material; and

c. You must comply with the conditions in Section 3(a) if You Share all or a substantial portion of the contents of the database.

For the avoidance of doubt, this Section 4 supplements and does not replace Your obligations under this Public License where the Licensed Rights include other Copyright and Similar Rights.

Section 5 -- Disclaimer of Warranties and Limitation of Liability.

a. UNLESS OTHERWISE SEPARATELY UNDERTAKEN BY THE LICENSOR, TO THE EXTENT POSSIBLE, THE LICENSOR OFFERS THE LICENSED MATERIAL AS-IS AND AS-AVAILABLE, AND MAKES NO REPRESENTATIONS OR WARRANTIES OF ANY KIND CONCERNING THE LICENSED MATERIAL, WHETHER EXPRESS, IMPLIED, STATUTORY, OR OTHER. THIS INCLUDES, WITHOUT LIMITATION, WARRANTIES OF TITLE, MERCHANTABILITY, FITNESS FOR A PARTICULAR PURPOSE, NON-INFRINGEMENT, ABSENCE OF LATENT OR OTHER DEFECTS, ACCURACY, OR THE PRESENCE OR ABSENCE OF ERRORS, WHETHER OR NOT KNOWN OR DISCOVERABLE. WHERE DISCLAIMERS OF WARRANTIES ARE NOT ALLOWED IN FULL OR IN PART, THIS DISCLAIMER MAY NOT APPLY TO YOU.

b. TO THE EXTENT POSSIBLE, IN NO EVENT WILL THE LICENSOR BE LIABLE TO YOU ON ANY LEGAL THEORY (INCLUDING, WITHOUT LIMITATION, NEGLIGENCE) OR OTHERWISE FOR ANY DIRECT, SPECIAL, INDIRECT, INCIDENTAL, CONSEQUENTIAL, PUNITIVE, EXEMPLARY, OR OTHER LOSSES, COSTS, EXPENSES, OR DAMAGES ARISING OUT OF THIS PUBLIC LICENSE OR USE OF THE LICENSED MATERIAL, EVEN IF THE LICENSOR HAS BEEN ADVISED OF THE POSSIBILITY OF SUCH LOSSES, COSTS, EXPENSES, OR DAMAGES. WHERE A LIMITATION OF LIABILITY IS NOT ALLOWED IN FULL OR IN PART, THIS LIMITATION MAY NOT APPLY TO YOU.

c. The disclaimer of warranties and limitation of liability provided above shall be interpreted in a manner that, to the extent possible, most closely approximates an absolute disclaimer and waiver of all liability.

Section 6 -- Term and Termination.

a. This Public License applies for the term of the Copyright and Similar Rights licensed here. However, if You fail to comply with this Public License, then Your rights under this Public License terminate automatically.

b. Where Your right to use the Licensed Material has terminated under Section 6(a), it reinstates:

1. automatically as of the date the violation is cured, provided it is cured within 30 days of Your discovery of the violation; or

2. upon express reinstatement by the Licensor.

For the avoidance of doubt, this Section 6(b) does not affect any right the Licensor may have to seek remedies for Your violations of this Public License.

c. For the avoidance of doubt, the Licensor may also offer the Licensed Material under separate terms or conditions or stop distributing the Licensed Material at any time; however, doing so will not terminate this Public License.

d. Sections 1, 5, 6, 7, and 8 survive termination of this Public License.

Section 7 -- Other Terms and Conditions.

a. The Licensor shall not be bound by any additional or different terms or conditions communicated by You unless expressly agreed.

b. Any arrangements, understandings, or agreements regarding the Licensed Material not stated herein are separate from and independent of the terms and conditions of this Public License.

Section 8 -- Interpretation.

a. For the avoidance of doubt, this Public License does not, and shall not be interpreted to, reduce, limit, restrict, or impose conditions on any use of the Licensed Material that could lawfully be made without permission under this Public License.

b. To the extent possible, if any provision of this Public License is deemed unenforceable, it shall be automatically reformed to the minimum extent necessary to make it enforceable. If the provision cannot be reformed, it shall be severed from this Public License without affecting the enforceability of the remaining terms and conditions.

c. No term or condition of this Public License will be waived and no failure to comply consented to unless expressly agreed to by the Licensor.

d. Nothing in this Public License constitutes or may be interpreted as a limitation upon, or waiver of, any privileges and immunities that apply to the Licensor or You, including from the legal processes of any jurisdiction or authority.


Creative Commons is not a party to its public licenses. Notwithstanding, Creative Commons may elect to apply one of its public licenses to material it publishes and in those instances will be considered the “Licensor.” The text of the Creative Commons public licenses is dedicated to the public domain under the CC0 Public Domain Dedication. Except for the limited purpose of indicating that material is shared under a Creative Commons public license or as otherwise permitted by the Creative Commons policies published at creativecommons.org/policies, Creative Commons does not authorize the use of the trademark "Creative Commons" or any other trademark or logo of Creative Commons without its prior written consent including, without limitation, in connection with any unauthorized modifications to any of its public licenses or any other arrangements, understandings, or agreements concerning use of licensed material. For the avoidance of doubt, this paragraph does not form part of the public licenses.

Creative Commons may be contacted at creativecommons.org.